Below are the assumptions made for the calculation of mammography room barrier thickness. A brief explanation of each assumption is included.
Parameter
Assumption
Explanation
Workload
1200 mA-min/wk
This is based on data from our inspection database. Some facilities average between 500-600 patients per month. Assume 600 patients per month, 4 weeks/month, 4 films/patient, 120 mAs/film
All occupiable areas are assumed for design purposes to be fully occupied. This is to allow for future changes in areas adjacent to the room
Primary use factor
0.25
For machines that are capable of permitting primary beams to miss the film holder, it is assumed that they are directed to any particular barrier 1/4 of the time
Secondary use factor
1
Secondary radiation always strikes the barrier
Leakage technique factor
35 kVp, 5 mA
Assume the machine will meet 100 mR/hr at 1 meter leakage limit when operated continuously at 35 kVp and 5 mA
Primary Distance
1.5 meters
Based on observation during inspections. Many machines are positioned close to the vertical barrier.
Secondary Distance
1.5 meters
Same
Leakage Distance
1.5 meters
Same
SID
65 cm
Most machines have SIDs near 65 cm
Beam size at SID
720 cm2
Area of a 24 by 30 cm cassette
Scattering Angle
135 degrees
A conservative scattering angle
Shielding design goal
0.02 mGy per week
(1 mGy per year)
Based on NCRP No. 147
Using the above assumptions for a mammography room, the barrier thickness was determined using XRAYBARR v 1.2 by D.J. Simpkin, Ph.D. These thicknesses are:
For machines manufactured after 9/30/99 (secondary radiation only)
mm
inches
Lead
0.139
5.5E-3
Concrete
16.6
0.65
Gypsum
45.8
1.8
Steel
0.731
0.029
Plate Glass
17.4
0.687
Wood
238
9.36
For machines manufactured on or prior to 9/30/99 (some primary radiation missing the film holder assembly is possible on some models if collimation misused)