PLEASE DISTRIBUTE THIS CHANGE TO ALL OF YOUR SERVICE PERSONNEL.
The following table serves as a replacement table for the one listed on page 3-28 for Section 3.31., "Vehicle-Tank Meters Code" of the 2003 Edition of NIST Handbook 44.
In particular, the tolerances for Accuracy Class 0.3 devices, under the column heading "Acceptance Tolerance," was changed from 0.2% to 0.15% and under the column heading "Special Test Tolerance," was changed from 0.5% to 0.45%.
In addition, a row was added for Accuracy Class 1.5 devices for Water applications.
These changes accurately reflect Item 331-3, "Tolerances, Table 1. Accuracy Classes for Vehicle-Tank Meters," of the Specifications and Tolerances Committee Final Report, adopted at the 87th National Conference on Weights and Measures in July 2002.
PDF
Version of this table (Requires Adobe
Acrobat)
|
Tolerances, Table 1.
Accuracy Classes for Vehicle-Tank Meters
|
|
Accuracy
Class
|
Application
|
Acceptance
Tolerance
|
Maintenance
Tolerance
|
Special Test
Tolerance
|
|
0.3
|
Petroleum products including large capacity motor fuel devices (flow rates over 115 L/min (30
gpm))**, heated products at or greater than 50 °C asphalt at or below temperatures 50
°C, all other liquids not shown where the typical delivery is over 200 L (50 gal) |
0.15 %
|
0.3 % |
0.45 % |
| 0.3A |
Asphalt at temperatures greater than 50
°C |
0.3 % |
0.3 % |
0.5 % |
| 0.5 |
Petroleum products delivered from small capacity (at 4 L/min (1
gpm) through 115 L/min (30 gpm))** motor-fuel devices, agri-chemical liquids, and all other applications not shown. |
0.3 % |
0.5 % |
0.5 % |
| 1.1 |
Petroleum products and other normal liquids from devices with flow rates** less than 1 gpm and devices designed to deliver less than one gallon. |
0.75 % |
1.0 % |
1.25 % |
| 1.5 |
Water |
Overregistration |
1.5 % |
1.5 % |
1.5 % |
| Underregistration |
1.5 % |
1.5 % |
5.0 % |
|
*The maintenance tolerances on normal and special tests for 5-gallon and 10-gallon test drafts are 6 cubic inches and 11
cubic inches, respectively. Acceptance tolerances on normal and special tests are 3 cubic inches and 5.5 cubic inches.
** Flow rate refers to designed or marked maximum flow rate.
|