Mobile Lithotripsy Shielding Guidelines

Mobile lithotripsy machines may be used for temporary clinical purposes in operating rooms with steel or leaded doors. Windows (in doors and walls) must be shielded to at least the same lead equivalency as the door. If a machine is to be used in an area that does not meet these criteria, approval of the radiation shielding must be obtained prior to clinical use in that location.

Lithotripter rooms which do not meet the above criteria (operating rooms with steel or leaded doors) are typically required to be shielded with the equivalent of at least 1/32-inch lead, including doors and windows. Floors and ceilings, if there is occupancy below or above, are normally required to also be equivalent to 1/32-inch lead or 2.5 inches of standard-density concrete (147 pounds per cubic foot).

Updated 2/21/2006



Related Documents
BHCS/HFS-852; Application for a Radiation Shielding Plan Review - 508021 bytes PDF icon
Related Content
 •  Dental Cephalometric and Tomographic Shielding Guidelines
 •  Dental CT Shielding Guidelines
 •  Podiatric X-Ray Shielding Guidelines
 •  Radiation Shielding Plan Review and General Information
 •  Radiographic or Radiographic-Fluoroscopic Shielding Guidelines
 •  CT Scanner Shielding Guidelines
 •  Linear Accelerator Shielding Guidelines
 •  Linear Accelerator Shielding Guidelines - Mazeless Vaults
 •  Mammography Shielding Guidelines
 •  Pain Management Mobile C-Arm Fluoroscopic Room Shielding Guidelines
 •  Shielding Design Guidance
QR code

Michigan.gov Home |  LARA Home |  State Web Sites |  Office of Regulatory Reinvention |  Spending & Accountability
Accessibility Policy |  Link Policy |  Privacy Policy |  Security Policy | Michigan News | Michigan.gov Survey


Copyright © 2001-2013 State of Michigan